Anaesthesia Machine MJ-560B1

FEATURES

Stable and reliable, meeting the basic needs of anesthesia surgery

5.7 "high-definition LCD display

High precision flow control system ensures precise regulation of oxygen

High precision Vaporizer, available in four types: Enflurane, Isoflurane, Sevoflurane, and Halothane

Highly integrated breathing circuit with APL valve, providing excellent sealing and safety

 

TECHNICAL PARAMETERS

PART OF ANESTHESIA MACHINE

■Main Body                

High-strength engineering plastic rack, light, beautiful and corrosion resistant

 

■Scope of Application:

Adult, Child

 

■Gas Source

O2: 0.28~0.6MPa

 

■Flow Meter

O2: 0.05~1.0L/min,1.1~10L/min

 

■Flow Rate of Rapid Oxygen Supply

25~75L/min


■Low Oxygen Pressure Alarm

There will be sound alarm when the oxygen pressure < 0.2MPa


 ■Vaporizer

It has the function of automatic compensation based on pressure, temperature, and flow rate. The regulation range of evaporator concentration is 0~5 vol%.

Among Halothane,Enflurane, Isoflurane and Sevoflurane, one can be chosen for application as required by the customer.


■Respiratory Circuit           

Working mode: All-close, Semi-close, Semi-open

APL: ≤12.5 kPa

 

■Respiratory Bellows          

Bellows for adults and children

Tidal volume range: 0~1500 ml

 

PART OF VENTILATOR

■Display Mode              

High-definition 5.7” LCD screen display


 ■Ventilation Mode          

IPPV,MANUAL

 

■Ventilation Parameters        

Tidal volume:50~1500 ml

Rate:4~80 bpm

I:E:2:1~1:4

Pressure range:1.0~6.0 kPa

 

■Parameters For Ventilation Monitoring

Tidal volume. Ventilation volume. IPPV rate.I/E. peak pressure of airway, pressure-time waveform

 

■Security Alarm System

Airway Pressure Alarm: Upper limit setting range: 1.0~6.0 kPa, Low limit setting range: 0.4~2.0 kPa

Per-minute Ventilation Volume Alarm: Upper limit setting range:1~20L/min,Low limit setting range: 0~19L/min

Tidal Volume alarm: Upper limit setting range: 0.5~2.0L,Low limit setting range   0.02~1.0L

Sustained High Pressure Alarm: It will give alarm when stress have consistently been higher than 25 cmH2O for 15s.

Suffocation Alarm: It will give alarm when stress have consistently been higher than       10 cmH2O and tidal volume is less than 120ml for 15s.

 

■Power Supply

AC 100~240V,50/60Hz