FEATURES
Stable and reliable, meeting the basic needs of anesthesia surgery
5.7 "high-definition LCD display
High precision flow control system ensures precise regulation of oxygen
High precision Vaporizer, available in four types: Enflurane, Isoflurane, Sevoflurane, and Halothane
Highly integrated breathing circuit with APL valve, providing excellent sealing and safety
TECHNICAL PARAMETERS
PART OF ANESTHESIA MACHINE
■Main Body
High-strength engineering plastic rack, light, beautiful and corrosion resistant
■Scope of Application:
Adult, Child
■Gas Source
O2: 0.28~0.6MPa
■Flow Meter
O2: 0.05~1.0L/min,1.1~10L/min
■Flow Rate of Rapid Oxygen Supply
25~75L/min
■Low Oxygen Pressure Alarm
There will be sound alarm when the oxygen pressure < 0.2MPa
■Vaporizer
It has the function of automatic compensation based on pressure, temperature, and flow rate. The regulation range of evaporator concentration is 0~5 vol%.
Among Halothane,Enflurane, Isoflurane and Sevoflurane, one can be chosen for application as required by the customer.
■Respiratory Circuit
Working mode: All-close, Semi-close, Semi-open
APL: ≤12.5 kPa
■Respiratory Bellows
Bellows for adults and children
Tidal volume range: 0~1500 ml
PART OF VENTILATOR
■Display Mode
High-definition 5.7” LCD screen display
■Ventilation Mode
IPPV,MANUAL
■Ventilation Parameters
Tidal volume:50~1500 ml
Rate:4~80 bpm
I:E:2:1~1:4
Pressure range:1.0~6.0 kPa
■Parameters For Ventilation Monitoring
Tidal volume. Ventilation volume. IPPV rate.I/E. peak pressure of airway, pressure-time waveform
■Security Alarm System
Airway Pressure Alarm: Upper limit setting range: 1.0~6.0 kPa, Low limit setting range: 0.4~2.0 kPa
Per-minute Ventilation Volume Alarm: Upper limit setting range:1~20L/min,Low limit setting range: 0~19L/min
Tidal Volume alarm: Upper limit setting range: 0.5~2.0L,Low limit setting range 0.02~1.0L
Sustained High Pressure Alarm: It will give alarm when stress have consistently been higher than 25 cmH2O for 15s.
Suffocation Alarm: It will give alarm when stress have consistently been higher than 10 cmH2O and tidal volume is less than 120ml for 15s.
■Power Supply
AC 100~240V,50/60Hz
Copyrights©2023 KOJUI MEDICAL All rights reserved